Background: We reviewed our experience with the Sorin Pericarbon (Sorin, Saluggia, Italy) valve implanted in the aortic position.
Methods: From January 1990 to January 1996, 143 consecutive patients had a Pericarbon valve implanted in the aortic position. The mean age was 75+/-5 years. Seventy-eight patients (55%) were in New York Heart Association (NYHA) class III or IV. Sixty patients (42%) had one or more concomitant procedures (51 coronary artery bypass grafting [CABG], 7 carotid endarterectomies, 9 others).
Results: The hospital mortality rate was 12% (17 of 143 patients). The follow-up was 100% complete and the median time was 42 months (range, 2 to 79 months). There were 36 late deaths, 20 being cardiac-related: 5 non-valve-related, 11 valve-related, and 4 sudden unexpected deaths. The 5-year actuarial survival was 57%+/-5%. There were 6 early valve failures related to a calcific stenosis at a median time of 36 months (range, 5 to 66 months). Three patients had to undergo another operation and one of these patients died. One patient died the day before the planned reoperation and 2 patients are followed with a symptomatic aortic stenosis but refuse reoperation. Freedom from structural deterioration was 93%+/-3% at 4 years. Echocardiographic examination was obtained in 73 patients at a median time of 42 months (range, 4 to 79 months). Four additional asymptomatic patients were found to have calcifications of their prosthesis. The 5-year freedom from thromboembolic events and from endocarditis were, respectively, 87%+/-5% and 92%+/-3%.
Conclusion: The surprisingly high rate of early failure due to calcific stenosis and of thromboembolic events of the Pericarbon valve implanted in the aortic position in the elderly made us discontinue its use in our institution.