Evaluation of a pulsatile pediatric ventricular assist device in an acute right heart failure model

Ann Thorac Surg. 1997 Nov;64(5):1374-80. doi: 10.1016/S0003-4975(97)00901-6.

Abstract

Background: The development of pulsatile ventricular assist devices for children has been limited mainly by size constraints. The purpose of this study was to evaluate the MEDOS trileaflet-valved, pulsatile, pediatric right ventricular assist device (stroke volume = 9 mL) in a neonatal lamb model of acute right ventricular failure.

Methods: Right ventricular failure was induced in ten 3-week-old lambs (8.6 kg) by right ventriculotomy and disruption of the tricuspid valve. Control group 1 (n = 5) had no mechanical support whereas experimental group 2 (n = 5) had right ventricular assist device support for 6 hours. The following hemodynamic parameters were measured in all animals: heart rate and right atrial, pulmonary arterial, left atrial, and systemic arterial pressures. Cardiac output was measured by an electromagnetic flow probe placed on the pulmonary artery.

Results: All results are expressed as mean +/- standard deviation and analyzed by Student's t test. A p value less than 0.05 was considered statistically significant. Base-line measurements were not significantly different between groups and included systemic arterial pressure, 80.6 +/- 12.7 mm Hg; right atrial pressure, 4.6 +/- 1.6 mm Hg; mean pulmonary arterial pressure, 15.6 +/- 4.2 mm Hg; left atrial pressure, 4.8 +/- 0.8 mm Hg; and cardiac output, 1.4 +/- 0.2 L/min. Right ventricular injury produced hemodynamics compatible with right ventricular failure in both groups: mean systemic arterial pressure, 38.8 +/- 10.4 mm Hg; right atrial pressure, 16.8 +/- 2.3 mm Hg; left atrial pressure, 1.4 +/- 0.5 mm Hg; and cardiac output, 0.6 +/- 0.1 L/min. All group 1 animals died at a mean of 71.4 +/- 9.4 minutes after the operation. All group 2 animals survived the duration of study. Hemodynamic parameters were recorded at 2, 4, and 6 hours on and off pump, and were significantly improved at all time points: mean systemic arterial pressure, 68.0 +/- 13.0 mm Hg; right atrial pressure, 8.2 +/- 2.3 mm Hg; left atrial pressure, 6.4 +/- 2.1 mm Hg; and cardiac output, 1.0 +/- 0.2 L/min.

Conclusions: The results demonstrate the successful creation of a right ventricular failure model and its salvage by a miniaturized, pulsatile right ventricular assist device. The small size of this device makes its use possible even in small neonates.

MeSH terms

  • Acute Disease
  • Animals
  • Blood Pressure
  • Cardiac Output
  • Cardiac Output, Low / physiopathology
  • Cardiac Output, Low / surgery*
  • Child
  • Equipment Design
  • Evaluation Studies as Topic
  • Heart Rate
  • Heart-Assist Devices*
  • Humans
  • Infant, Newborn
  • Sheep
  • Ventricular Dysfunction, Right / physiopathology
  • Ventricular Dysfunction, Right / surgery*