Improvement of spiral MRI with the measured k-space trajectory

J Magn Reson Imaging. 1997 Sep-Oct;7(5):938-40. doi: 10.1002/jmri.1880070525.

Abstract

The k-space trajectory of a spiral imaging sequence was measured with a self-encoding technique. The image quality improved dramatically when reconstructed with the measured k-space trajectory. There were substantial artifacts in images reconstructed with the derived k-space trajectory under the assumption of gradient system linearity. The results indicated the non-linearity of the gradient system and the effectiveness of the correction technique.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Analog-Digital Conversion
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Models, Theoretical
  • Phantoms, Imaging
  • Sensitivity and Specificity