Test/retest reproducibility of iodine-123-betaCIT SPECT brain measurement of dopamine transporters in Parkinson's patients

J Nucl Med. 1997 Sep;38(9):1453-9.

Abstract

Iodine-123-beta-CIT has been used as a probe of dopamine transporters in Parkinson's disease patients using SPECT. We studied the test/retest reproducibility of SPECT measures in Parkinson's disease patients and healthy controls obtained after injection of [123I])beta-CIT in part to assess the utility of this tracer for longitudinal evaluation of striatal dopamine transporters as a marker of disease progression.

Methods: Seven Parkinson's disease patients and seven healthy control subjects participated in two [123I]beta-CIT SPECT scans separated by 7-21 days. Subjects were imaged at 24 hr post injection of 360 MBq (9.7 mCi) of [123I]beta-CIT. Two outcome measures were evaluated; 1) the ratio of specific striatal (activity associated with DA transporter binding) to nondisplaceable uptake, also designated V3," and 2) the total specific striatal uptake (%SSU) expressed as a percentage of injected radiotracer dose. For both measures, test/retest variability was calculated as the absolute difference of test minus retest divided by the mean of test/retest and expressed as a percent. In addition, the reproducibility of left and right striatal asymmetry and putamen:caudate ratios were determined.

Results: The two outcome measures demonstrated excellent test/retest reproducibility for both the Parkinson's disease and healthy subject groups with variability of striatal V3" = 16.8 +/- 13.3% and percent striatal uptake = 6.8 +/- 3.4% for Parkinson's disease patients and V3" = 12.8 +/- 8.9% and %SSU = 7.0 +/- 3.9% for control subjects. There were no statistically significant differences in test/retest variability between control subjects and Parkinson's disease patients for either outcome measure. The reproducibility of left/right asymmetry indices and putamen-to-caudate ratios showed no patient versus control subject differences. The asymmetry index had greater test/retest variability than the other outcome measures.

Conclusion: These data suggest that SPECT imaging performed at 24 hr postinjection of [123I]beta-CIT permits calculation of reliable and reproducible measures of dopamine transporters in both Parkinson's disease patients and control subjects and supports the feasibility of using [123I]beta-CIT in the evaluation of disease progression in Parkinson's disease.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Aged
  • Brain / diagnostic imaging
  • Brain / metabolism*
  • Carrier Proteins / analysis*
  • Cocaine / analogs & derivatives*
  • Corpus Striatum / metabolism
  • Dopamine / metabolism*
  • Dopamine Plasma Membrane Transport Proteins
  • Female
  • Humans
  • Iodine Radioisotopes*
  • Male
  • Membrane Glycoproteins*
  • Membrane Transport Proteins*
  • Middle Aged
  • Nerve Tissue Proteins*
  • Parkinson Disease / diagnostic imaging
  • Parkinson Disease / metabolism*
  • Putamen / metabolism
  • Reproducibility of Results
  • Tomography, Emission-Computed, Single-Photon*

Substances

  • Carrier Proteins
  • Dopamine Plasma Membrane Transport Proteins
  • Iodine Radioisotopes
  • Membrane Glycoproteins
  • Membrane Transport Proteins
  • Nerve Tissue Proteins
  • 2beta-carbomethoxy-3beta-(4-iodophenyl)tropane
  • Cocaine
  • Dopamine