Quantitative assessment of the presence of a single leg separation in Björk-Shiley convexoconcave prosthetic heart valves

Invest Radiol. 1997 Sep;32(9):540-9. doi: 10.1097/00004424-199709000-00006.

Abstract

Rationale and objectives: The authors developed an analytic software package for the objective and reproducible assessment of a single leg separation (SLS) in the outlet strut of Björk-Shiley convexoconcave (BSCC) prosthetic heart valves.

Methods: The radiographic cinefilm recordings of 18 phantom valves (12 intact and 6 SLS) and of 43 patient valves were acquired. After digitization of regions of interest in a cineframe, several processing steps were carried out to obtain a one-dimensional corrected and averaged density profile along the central axis of each strut leg. To characterize the degree of possible separation, two quantitative measures were introduced: the normalized pit depth (NPD) and the depth-sigma ratio (DSR). The group of 43 patient studies was divided into a learning set (25 patients) and a test set (18 patients).

Results: All phantom valves with an SLS were detected (sensitivity, 100%) at a specificity of 100%. The threshold values for the NPD and the DSR to decide whether a fracture was present or not were 3.6 and 2.5, respectively. On the basis of the visual interpretations of the 25 patient studies (learning set) by an expert panel, it was concluded that none of the patients had an SLS. To achieve a 100% specificity by quantitative analysis, the threshold values for the NPD and the DSR were set at 5.8 and 2.5, respectively, for the patient data. Based on these threshold values, the analysis of patient data from the test set resulted in one false-negative detection and three false-positive detections.

Conclusions: An analytic software package for the detection of an SLS was developed. Phantom data showed excellent sensitivity (100%) and specificity (100%). Further research and software development is needed to increase the sensitivity and specificity for patient data.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Electronic Data Processing
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Prosthesis Failure
  • Radiographic Image Enhancement / methods*
  • Sensitivity and Specificity