Differences in phycoerythrin- or fluorescein-isothiocyanate conjugated 8G12 on CD34+ cell evaluation

Haematologica. 1997 May-Jun;82(3):334-5.

Abstract

To evaluate the equivalence of 8G12 conjugated with either phycoerythrin (PE) or fluorescein isothiocyanate (FITC), duplicate fluorochrome-8G12 labelling was carried out in 229 samples. Additionally, 53 samples were simultaneously immunostained with FITC-8G12 and PE-8G12. Significantly higher values (p < 0.001) were observed in PE-CD34+ cells when compared with FITC-CD34+ cells both in duplicate and simultaneous analysis. Our data suggest that the choice of fluorochrome is relevant in the measurement of CD34+ cells with 8G12.