Evaluation of a second-generation thyrotropin automated immunoassay

Eur J Clin Chem Clin Biochem. 1996 Jan;34(1):73-6.

Abstract

We evaluated the analytical performance of the Vista automated immunoassay system for human thyrotropin determination. The operating characteristics as well as the analytical performance were assessed. The Vista human thyrotropin immunoassay showed a minimal detection limit of 0.08 mU/l and a functional sensitivity of 0.12 mU/l. The system meets the criteria for second-generation human thyrotropin assays.

MeSH terms

  • Automation
  • Evaluation Studies as Topic
  • Humans
  • Immunoenzyme Techniques*
  • Reagent Kits, Diagnostic*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thyrotropin / analysis*

Substances

  • Reagent Kits, Diagnostic
  • Thyrotropin