Clinical and analytical evaluation of a continuous enzymatic method for measuring pancreatic lipase activity

Clin Chem. 1993 Feb;39(2):304-8.

Abstract

We report the evaluation of a new commercial kit for the determination of pancreatic lipase activity. The kit is based on the use of a 1,2-diglyceride as substrate and a specific monoglyceride lipase. The detection step is the continuous colorimetric measurement of hydrogen peroxide produced from glycerol by glycerol kinase, glycerol-3-phosphate oxidase, and peroxidase reactions. The procedure appears to be precise (between-day CV < 9%) and the results show good correlation with those obtained by alternative procedures (vs turbidimetry, r = 0.965; vs ultraviolet absorbance-enzymatic method, r = 0.995; vs Ektachem, r = 0.976; vs immunometry, r = 0.970). However, the method is susceptible to interference by increased concentrations (> 4.5 mmol/L) of serum triglycerides. We estimated the reference interval for healthy adults to be 8-44 U/L. When we evaluated clinical efficacy by using receiver-operating characteristic curves and the overlap index, no significant differences were found between the commercial kit and a common turbidimetric assay for diagnosing patients with acute pancreatitis; both methods performed satisfactorily.

MeSH terms

  • Adult
  • Aged
  • Colorimetry
  • Diglycerides / metabolism
  • Evaluation Studies as Topic
  • Female
  • Glycerol Kinase
  • Glycerolphosphate Dehydrogenase
  • Humans
  • Hydrogen Peroxide / analysis
  • Indicators and Reagents
  • Infant
  • Lipase / blood*
  • Male
  • Middle Aged
  • Nephelometry and Turbidimetry
  • Pancreas / enzymology*
  • Pancreatitis / enzymology
  • Peroxidase
  • Reagent Kits, Diagnostic / standards*
  • Reagent Kits, Diagnostic / statistics & numerical data
  • Reference Values

Substances

  • Diglycerides
  • Indicators and Reagents
  • Reagent Kits, Diagnostic
  • Hydrogen Peroxide
  • Glycerolphosphate Dehydrogenase
  • Peroxidase
  • Glycerol Kinase
  • Lipase