The results of a receiver operator characteristic (ROC) study comparing maximum likelihood estimator (MLE) reconstructions of human FDG PET brain scan data to filtered backprojection reconstructions of the same data are reported. The purpose of the study was to determine whether MLE reconstructions would result in higher detectability of small focal lesions introduced artificially into otherwise normal scan data. One physician assisted in defining the location and intensity of the lesions and five physicians read the final images. Data from 90 datasets were used for the study. Of those, 42 were left in their original "normal" condition and 48 were modified by added lesions. All datasets were reconstructed by the two methods and submitted to the five physicians for evaluation. The results show an increase in the area under the ROC curve from approximately 0.65 for filtered backprojection to approximately 0.71 for the maximum likelihood reconstructions for four of the five observers with good statistical significance.