Optimization of MR-ARFI for Human Transcranial Focused Ultrasound

bioRxiv [Preprint]. 2024 Nov 15:2024.11.13.623314. doi: 10.1101/2024.11.13.623314.

Abstract

Magnetic resonance acoustic radiation force imaging (MR-ARFI) is an exceptionally promising technique to non-invasively confirm targeting accuracy and estimate exposure of low-intensity transcranial focused ultrasound stimulation. MR-ARFI uses magnetic field motion encoding gradients to visualize the MR phase changes generated by microscopic displacements at the ultrasound focus. Implementing MR-ARFI in the human central nervous system has been hindered by 1) phase distortion caused by subject motion, and 2) insufficient signal-to-noise ratio at low (<1.0 MPa) ultrasound pressures. The purpose of this study was to optimize human MR-ARFI to allow reduced ultrasound exposure while at the same time being robust to bulk and physiological motion. We demonstrate that a time series of single-shot spiral acquisitions, while triggering ultrasound on and off in blocks, provides ARFI maps that with correction are largely immune to bulk and pulsatile brain motion. Furthermore, the time series approach allows for a reduction in ultrasound exposure per slice while improving motion robustness with reduced scan time. The focused ultrasound beam can be visualized in an 80 second scan with our protocol, enabling iteration for image-guided targeting. We demonstrated robust ARFI signals at the expected target in 4 participants. Our results provide persuasive proof-of-principle that MR-ARFI can be used as a tool to guide ultrasound-based precision neural circuit therapeutics.

Publication types