Functional Ventricular Undersensing After Ventricular Ectopic Beats Due to ICD Dynamic Sensitivity

Pacing Clin Electrophysiol. 2025 Jan;48(1):50-52. doi: 10.1111/pace.15099. Epub 2024 Oct 24.

Abstract

An 80-year-old patient, previously implanted for idiopathic sustained ventricular tachycardia with a Rivacor VR-T DX implantable cardioverter defibrillator (ICD Biotronik, Berlin, Germany), came to our pacemaker clinic for palpitations. Electrocardiogram (ECG) showed sinus bradycardia with interpolated ventricular ectopic beats. A ventricular-paced event occurred with a coupling interval shorter than the expected one suggesting ventricular undersensing. The phenomenon can be explained by the default ICD sensitivity setting proposed by the manufacturer.

Keywords: ICD; sensitivity; undersensing.