Feasibility of thin-slice, low noise images created using multi-kernel synthesis to replace multiple image series in head CT

Acta Radiol. 2024 Oct 17:2841851241280365. doi: 10.1177/02841851241280365. Online ahead of print.

Abstract

Background: SynthesiZed Improved Resolution and Concurrent nOise reductioN (ZIRCON) is a multi-kernel synthesis method that creates a single series of thin-slice computed tomography (CT) images displaying low noise and high spatial resolution, increasing reader efficiency and minimizing partial volume averaging.

Purpose: To compare the diagnostic performance of a single set of ZIRCON images to two routine clinical image series using conventional CT head and bone reconstruction kernels for diagnosing intracranial findings and fractures in patients with trauma or suspected acute neurologic deficit.

Material and methods: In total, 50 patients underwent clinically indicated head CT in the ER (15 normal, 35 abnormal cases). A non-reader neuroradiologist established the reference standard. Three neuroradiologists reviewed two routine clinical series (head and bone kernels) and a single ZIRCON series, detecting intracranial findings or fractures and rating confidence (0-100). Sensitivity, specificity, and jackknife free-response receiver operating characteristic (JAFROC) figure of merit (FOM) were compared (limit of non-inferiority: -0.10).

Results: ZIRCON and conventional images demonstrated comparable performance for fractures (sensitivity: 51.5% vs. 54.5%; specificity: 40.2% vs. 34.2%) and intracranial findings (sensitivity: 88.2% vs. 91.4%; specificity: 77.2% vs. 73.7%).The estimated difference of JAFROC FOM demonstrated ZIRCON non-inferiority for acute pathologies overall (0.003 [95% CI=-0.051-0.057]) and fractures (0.048 [95% CI=-0.050-0.145]) but not for intracranial findings alone (-0.024 [95% CI=-0.100-0.052]).

Conclusion: Thin-slice, low noise, and high spatial resolution images can be created to display intracranial findings and fractures replacing multiple images series in head CT with similar performance. Future studies in more patients and further algorithmic development are warranted.

Keywords: Computed tomography; hemorrhage; image postprocessing; trauma.