Standardization of scan protocols for RT CT simulator from different vendors using quantitative image quality technique

J Appl Clin Med Phys. 2024 Oct;25(10):e14484. doi: 10.1002/acm2.14484. Epub 2024 Aug 13.

Abstract

Objective: To investigate the feasibility of standardizing RT simulation CT scanner protocols between vendors using target-based image quality (IQ) metrics.

Method and materials: A systematic assessment process in phantom was developed to standardize clinical scan protocols for scanners from different vendors following these steps: (a) images were acquired by varying CTDIvol and using an iterative reconstruction (IR) method (IR: iDose and model-based iterative reconstruction [IMR] of CTp-Philips Big Bore scanner, SAFIRE of CTs-Siemens biograph PETCT scanner), (b) CT exams were classified into body and brain protocols, (c) the rescaled noise power spectrum (NPS) was calculated, (d) quantified the IQ change due to varied CTDIvol and IR, and (e) matched the IR strength level. IQ metrics included noise and texture from NPS, contrast, and contrast-to-noise ratio (CNR), low contrast detectability (d'). Area under curve (AUC) of the receiver operation characteristic curve of d' was calculated and compared.

Results: The level of change in the IQ ratio was significant (>0.6) when using IMR. The IQ ratio change was relatively low to moderate when using either iDose in CTp (0.1-0.5) or SAFIRE in CTs (0.1-0.6). SAFIRE-2 in CTs showed a closer match to the reference body protocol when compared to iDose-3 in CTp. In the brain protocol, iDose-3 in CTp could be matched to the low to moderate level of SAFIRE in CTs. The AUC of d' was highest when using IMR in CTp with lower CTDIvol, and SAFIRE in CTs performed better than iDose in CTp CONCLUSION: It is possible to use target-based IQ metrics to evaluate the performance of the system and operations across various scanners in a phantom. This can serve as an initial reference to convert clinical scanned protocols from one CT simulation scanner to another.

Keywords: NPS; TTF; detectability; iterative reconstruction.

MeSH terms

  • Algorithms
  • Humans
  • Image Processing, Computer-Assisted* / methods
  • Neoplasms / diagnostic imaging
  • Neoplasms / radiotherapy
  • Phantoms, Imaging*
  • Radiographic Image Interpretation, Computer-Assisted / methods
  • Radiographic Image Interpretation, Computer-Assisted / standards
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy, Intensity-Modulated / methods
  • Signal-To-Noise Ratio
  • Tomography Scanners, X-Ray Computed / standards
  • Tomography, X-Ray Computed* / instrumentation
  • Tomography, X-Ray Computed* / methods
  • Tomography, X-Ray Computed* / standards