Implantable loop recorder uncovered torsades de pointes in long-QT syndrome type 1 with multi cause of syncope

J Cardiol Cases. 2024 Feb 2;29(5):214-217. doi: 10.1016/j.jccase.2024.01.006. eCollection 2024 May.

Abstract

An implantable loop recorder (ILR) is now widely used for differential diagnosis of unexplained syncope or recurrent syncope with unknown causes. In the inherited arrhythmia syndromes, ILR may be useful for management of the therapeutic strategies; however, there is no obvious evidence to uncover arrhythmic syncope by ILR in long-QT syndrome (LQTS) patients. Here we experienced a 19-year-old female patient with LQTS type 1 who had recurrent syncope even after beta-blocker therapy but no arrhythmias were documented, and some episodes might be due to non-cardiogenic causes. Implantable cardioverter defibrillator (ICD) therapy was also recommended; however, she could not accept ICD but was implanted with ILR for further continuous monitoring. Two years later, she suffered syncope during a brief run, and ILR recorded an electrocardiogram at that moment. Thus a marked QT interval prolongation as well as T-wave alternance resulting in development of torsades de pointes could be detected. Although ILR is just a diagnostic tool but does not prevent sudden cardiac death, most arrhythmic events in LQTS are transient and sometimes hard to be diagnosed as arrhythmic syncope. ILR may provide direct supportive evidence to select the optimal therapeutic strategy in cases where syncope is difficult to diagnose.

Learning objective: Long-QT syndrome (LQTS) patients often suffer recurrent syncope even after beta-blocker therapy, but torsades de pointes (TdP) is not always detected by standard 12‑lead electrocardiogram or Holter monitoring, and some syncope might be non-cardiogenic. In this case, implantable loop recorder (ILR) documented the evidence of QT interval prolongation and beat-by-beat T-wave alternance subsequent TdP. Thus, ILR may provide useful evidence for the optimal treatment strategy in LQTS cases where syncope is difficult to diagnose.

Keywords: Diagnosis; Implantable loop recorder; Long-QT syndrome; Syncope; T-wave alternance; Torsades de pointes.

Publication types

  • Case Reports