Epicardial pacing outcomes in infants with heart block: Lead and device complications from a multicenter experience

Heart Rhythm. 2025 Jan;22(1):170-180. doi: 10.1016/j.hrthm.2024.07.014. Epub 2024 Jul 14.

Abstract

Background: Infants with complete heart block (CHB) require epicardial pacemaker (PM) insertion. Prior studies described epicardial pacing outcomes in infants and children, although they were limited by small or heterogeneous populations.

Objective: This study aimed to explore patient- and procedure-level associations with device complications in infants with CHB who received a permanent PM.

Methods: This was a multicenter, retrospective cohort study including infants receiving an epicardial PM between 2000 and 2021 for CHB. The primary outcome was time to device-related adverse event: lead failure requiring revision; pocket infection; exit block requiring increased pacing output; or lead-related coronary artery compression. Time-to-event analysis was performed by the Kaplan-Meier method with a multivariable Cox proportional hazards model.

Results: There were 174 infants who received an epicardial PM (282 bipolar, 39 unipolar leads) for CHB. Median age and weight at PM were 93.5 days and 4.5 kg, respectively. Pacing indication was postoperative CHB in 63% and congenital CHB in 37%. The median follow-up was 2.1 years. The primary outcome occurred in 26 infants at a median time to event of 0.6 year. Age ≤90 days at PM implantation was the most significant risk factor for a device-related adverse event (hazard ratio, 7.02; P < .001), primarily driven by pocket infections. Lead failure occurred in 3% of leads with a 5- and 10-year freedom from failure of 93% and 83%, respectively.

Conclusion: Device complications affect 15% of infants receiving a permanent PM for heart block. Age ≤90 days at PM implantation is especially associated with infectious complications. Epicardial lead durability appears similar to previously reported pediatric experiences.

Keywords: Complication; Heart block; Infant; Lead; Pacemaker; Pediatric.

Publication types

  • Multicenter Study

MeSH terms

  • Cardiac Pacing, Artificial / adverse effects
  • Cardiac Pacing, Artificial / methods
  • Equipment Failure
  • Female
  • Follow-Up Studies
  • Heart Block* / etiology
  • Heart Block* / therapy
  • Humans
  • Infant
  • Infant, Newborn
  • Male
  • Pacemaker, Artificial* / adverse effects
  • Pericardium
  • Retrospective Studies
  • Risk Factors
  • Treatment Outcome