Bicuspid valve CT registry of balloon-expandable TAVR: BETTER TAVR registry

Catheter Cardiovasc Interv. 2024 Jul;104(1):105-114. doi: 10.1002/ccd.31091. Epub 2024 May 31.

Abstract

Background: The anatomic substrate of bicuspid valves may lead to suboptimal TAVR stent expansion and geometry.

Aim: We evaluated determinants of stent geometry in bicuspid valves treated with Sapien transcatheter aortic valve replacement (TAVR) valves.

Methods: A multicenter retrospective registry of patients (February 2019 to August 2022) who underwent post-TAVR computed tomography to determine stent area (vs. nominal valve area) and stent ellipticity (maximum diameter/minimum diameter). Predictors of relative stent expansion (minimum area/average of inflow + outflow area) and stent ellipticity were evaluated in a multivariable regression model, including valve calcium volume (indexed by annular area), presence of raphe calcium, sinus diameters indexed by area-derived annular diameter, and performance of pre-dilation and post-dilation.

Results: The registry enrolled 101 patients from four centers. The minimum stent area (vs. nominal area) was 88.1%, and the maximum ellipticity was 1.10, with both observed near the midframe of the valve in all cases. Relative stent expansion ≥90% was observed in 64/101 patients. The only significant predictor of relative stent expansion ≥90% was the performance of post-dilation (OR: 4.79, p = 0.018). Relative stent expansion ≥90% was seen in 86% of patients with post-dilation compared to 57% without (p < 0.001). The stent ellipticity ≥1.1 was observed in 47/101 patients. The significant predictors of stent ellipticity ≥1.1 were the indexed maximum sinus diameter (OR: 0.582, p = 0.021) and indexed intercommisural diameter at 4 mm (OR: 2.42, p = 0.001). Stent expansion has a weak negative correlation with post-TAVR mean gradient (r = -0.324, p < 0.001).

Conclusion: Relative stent expansion ≥90% was associated with the performance of post-dilation, and stent ellipticity ≥1.1 was associated with indexed intercommisural diameter and indexed maximum sinus diameter. Further studies to determine optimal deployment strategies in bicuspid valves are needed.

Keywords: TAVR; balloon expandable valve; bicuspid valve; computed tomography.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / physiopathology
  • Aortic Valve Stenosis* / surgery
  • Aortic Valve* / diagnostic imaging
  • Aortic Valve* / physiopathology
  • Aortic Valve* / surgery
  • Balloon Valvuloplasty / adverse effects
  • Bicuspid Aortic Valve Disease / diagnostic imaging
  • Bicuspid Aortic Valve Disease / surgery
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Prosthesis Design*
  • Registries*
  • Retrospective Studies
  • Risk Factors
  • Stents
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Transcatheter Aortic Valve Replacement* / instrumentation
  • Treatment Outcome
  • United States