Background: Failure due to trunnionosis with adverse local tissue reaction (ALTR) has been reported with cobalt-chrome (CoCr) heads in total hip arthroplasty (THA); however, there are limited data on the use of these heads in the revision setting. The purpose of this study was to analyze the outcomes of patients who underwent revision THA with a retained femoral component and received a CoCr femoral head on a used trunnion.
Methods: In this retrospective review, we identified all patients who underwent revision THA with a retained femoral component and received a CoCr femoral head between February 2006 and March 2014. Demographic factors, implant details, and postoperative complications, including the need for repeat revisions, were recorded. In total, 107 patients were included (mean age 67 years, 74.0% women). Of the 107 patients, 24 (22.4%) required repeat revisions.
Results: Patients who required repeat revision were younger than those who did not (mean age: 62.9 versus 69, P = .03). The most common indications for repeat revision were instability (8 of 24, 33.3%), ALTR (5 of 24, 20.8%), and infection (4 of 18, 16.7%). Evidence of ALTR or metallosis was identified at the time of reoperation in 10 of the 24 patients who underwent re-revision (41.7%).
Conclusions: The placement of a new CoCr femoral head on a used trunnion during revision THA with a retained femoral component carries a significant risk of complication (22.4%) and should be avoided when possible.
Keywords: cobalt-chrome; metallosis; revision; total hip arthroplasty; trunnion.
Copyright © 2024 Elsevier Inc. All rights reserved.