Background: Optimization of automated insulin delivery (AID) settings is required to achieve desirable glycemic outcomes. We evaluated safety and efficacy of a computerized system to initialize and adjust insulin delivery settings for the t:slim X2 insulin pump with Control-IQ technology in adults with type 1 diabetes (T1D).
Methods: After a 2-week continuous glucose monitoring (CGM) run-in period, adults with T1D using multiple daily injections (MDI) (N = 33, mean age 36.1 years, 57.6% female, diabetes duration 19.7 years) were transitioned to 13 weeks of Control-IQ technology usage. A computerized algorithm generated recommendations for initial pump settings (basal rate, insulin-to-carbohydrate ratio, and correction factor) and weekly follow-up settings to optimize glycemic outcomes. Physicians could override the automated settings changes for safety concerns.
Results: Time in range 70 to 180 mg/dL improved from 45.7% during run-in to 69.1% during the last 30 days of Control-IQ use, a median improvement of 18.8% (95% confidence interval [CI]: 13.6-23.9, P < .001). This improvement was evident early in the study and was sustained over 13 weeks. Time <70 mg/dL showed a gradual decreasing trend over time. Percentage of participants achieving HbA1c <7% went from zero at baseline to 55% at study end (P < .001). Only six of the 318 automated settings adaptations (1.9%) were overridden by study investigators.
Conclusions: Computerized initiation and adaptation of Control-IQ technology settings from baseline MDI therapy was safe in adults with T1D. The use of this simplified system for onboarding and optimizing Control-IQ technology may be useful to increase uptake of AID and reduce staff and patient burden in clinical care.
Keywords: Control-IQ; adaptation; automated insulin delivery; initialization; t:slim X2.