Hisian Lead Electrical Decay After TAVR Intervention

JACC Case Rep. 2023 Nov 8:28:102090. doi: 10.1016/j.jaccas.2023.102090. eCollection 2023 Dec 20.

Abstract

We report a case of patient with a single-chamber Hisian pacemaker who developed complete atrioventricular block and significant deterioration of the ventricular threshold and sensing after transcatheter aortic valve replacement. Revision of the implantation was required, with ventricular lead extraction and replacement.

Keywords: Hisian pacemaker; aortic stenosis; atrioventricular block; computed tomography; transcatheter aortic valve implantation; transcatheter heart valve; virtual basal ring.

Publication types

  • Case Reports