Background: Direct conversion x-ray Photon Counting Detectors (PCD) are posed to play a vital role in future medical imaging devices such as Computed Tomography (CT) scanners. PCD are expected to improve current CT technology on several fronts, such as resolution, dose utilization, and spectral performance. However, they are not readily expected to improve the handling of object scatter, one of the major sources of image artifacts in CT technology.
Purpose: We explore a potential method for obtaining in-situ object scatter estimation using the same PCD array used in the x-ray imaging system, such as in computed tomography. This unexpected benefit of using PCD has the potential to improve the image quality by providing better input into the scatter estimation and correction algorithms used in image reconstruction.
Methods: In CT scanners the primary method for rejecting scatter signal originating from the scanned object relies on placing anti-scatter grids (ASG) close to the detector plane. This remains the case when transitioning to using PCD instead of energy integration detectors in CT. However, the combination of PCD and ASG opens a possibility to use some of the unique properties of PCD, namely, very low noise and coincidence counters to obtain, in addition to the attenuation data, a simultaneous and instantaneous estimate of the scatter signal reaching every detector element. When a small air gap is introduced between the ASG and the detector surface, the scatter radiation with large angular distribution has a greater probability of producing charge sharing events that can be detected by a coincidence counter. In this work we demonstrate the feasibility of such an approach in a tabletop experiment using PCD detector that lacks coincidence counting capability, instead we use the spectral signature of split charge events as proxy to coincidence counting. For this purpose, we first demonstrate the spectral impact of ASG misalignment using the same experimental setup. In addition, we perform a separate tabletop scattering experiment from a narrow column of water that demonstrates another potential use of the low noise capabilities of PCDs.
Results: We measured and quantified the high sensitivity of the spectral response to ASG alignment on the PCD detector pixel array, we found that the probability of energy misregistration of 60 keV photons can increase by up to a factor of 3 when the ASG is poorly aligned. We then leveraged these results to obtain an estimate on the expected increase in coincidence counts for a wide range of scatter-to-primary (SPR) ratio and find a good match with expectations from a geometric modeling of the system, where the expected increase in coincidences was of the order of the SPR. Finally, the low noise detector also allowed us to measure the real space scatter signal associated with the coherent molecular form factor of water, revealing the ring-shaped scatter signal with an energy dependent distribution that was well captured by calculation.
Conclusions: The advent of PCD detectors and their imminent use in commercial CT scanners opens new and exciting possibilities for utilizing PCD detectors in unexpected ways. In this proof-of-concept study, we showed how charge sharing, a spectral information degrading effect, can instead be used to obtain in-situ scatter estimation. We also demonstrated the PCD ability to perform extremely sensitive measurements using affordable benchtop setup for investigations normally reserved for synchrotron facilities.
Keywords: photon counting; scatter.
© 2023 American Association of Physicists in Medicine.