With the incidence of breast cancer rising to the top among female malignant tumors, magnetic resonance images guided breast biopsy intervention and minimally invasive treatment have developed as a clinically practical research issue. High field studies have shown the diagnostic value of breast MRI, but the examination costs greatly exceed those of competing conventional mammography. In this case, low-field MRI cannot merely provide typical MRI contrast, but also significantly reduce the cost of diagnosis and treatment for breast cancer patients. This work describes a unilateral breast coil and prototype intervention device, which provides a customized solution for low-field MRI-guided breast intervention. Results demonstrate that the low-field MRI breast intervention device facilitates medical intervention procedures. And the designed positioning device can locate the target lesion within 2-3 mm accuracy. Phantom tests with the customized unilateral coil indicate that the open loops perform as well as the 4-channel commercial closed breast coil, presenting a relatively good SNR (signal-to-noise ratio) and uniformity characteristics. MR scanning images of the volunteer breast using the breast intervention coil also show high SNR, which lays a foundation for further implementation of image-guided breast interventional minimally invasive surgery with the low-field MRI system.
Keywords: Breast MRI; Intervention device; Lesion localization; Low-field MRI; Open coil; SNR maps; Uniformity.
Copyright © 2023. Published by Elsevier Inc.