During endoscopic ultrasound-guided biliary drainage (EUS-BD), there is a risk for bile leakage until stent deployment, which can result in severe peritonitis, particularly when passing a drainage stent becomes challenging despite tract dilation. There is no established method or dedicated device to optimize EUS-BD. Therefore, we have developed a novel stent deployment technique using the tapered sheath dilator. To address the safety and technical aspects of the EUS-BD technique, we retrospectively analyzed 11 consecutive patients who underwent EUS-BD using the tapered sheath dilator. The procedure involved the insertion of a guidewire, followed by mechanical dilation using the tapered sheath dilator. Subsequently, the inner catheter was removed and drainage stents (up to 6 Fr in diameter) were deployed through the outer sheath. We found a 100% technical success rate for tract dilation and stent deployment; moreover, all patients achieved clinical success. The median time for dilation was 40 s (range, 8-198), whereas the median time from dilation to stent deployment was 10 min (range, 6-19). Notably, no cases of bile leakage or peritonitis were observed. In conclusion, the use of the integrated device for tract dilation and stent delivery system might provide a safe and straightforward technique for drainage stenting during EUS-BD.
Keywords: EUS‐BD; Endosheather; bile leakage; biliary drainage; tapered sheath dilator.