Establishing quality control action limits for CT number accuracy in spectral images using an American College of Radiology phantom

Med Phys. 2023 Oct;50(10):6071-6078. doi: 10.1002/mp.16626. Epub 2023 Jul 20.

Abstract

Background: Computed tomography (CT) number accuracy is important for quality assurance in CT imaging. However, in dual-energy CT imaging, there are no widely used action limits for CT number accuracy in spectral images, information that is urgently needed.

Purpose: To establish action limits for spectral CT images using longitudinal spectral data and an American College of Radiology (ACR) phantom.

Methods: An ACR accreditation phantom was scanned routinely as part of a quality control program in our institution. We selected and analyzed 57 continuous weekly scans. The CT numbers or the density values of conventional and spectral images, including virtual monoenergetic images (40, 50, 70, 120, and 200 keV), iodine maps, calcium suppressed, and virtual non-contrast images, were measured in the four inserts (solid water, bone, polyethylene, and acrylic) of the phantom. Longitudinal data were analyzed for correlation using Pearson's correlation coefficient (r) and standard deviation (SD). The SD ratios between spectral images and conventional images were calculated and the action limits for spectral images were established based on the action limits from the ACR.

Results: Strong to very strong correlations (r > 0.70 or r < -0.70) were found among most spectral image types except the 200 keV images using solid water, polyethylene, and acrylic inserts (r = [-0.45, 0.64]). The SD ratio was highest for the 40 keV images, ranging from 2.8 to 6.5. The action limits of the bone insert were baseline ± 5.3 mg/mL for the iodine map and ranged from baseline ± 23.0 HU to baseline ± 391.9 HU for the other image types. The action limits for solid water ranged from baseline ± 4.1 HU to baseline ± 25.3 HU. The results for the polyethylene and the acrylic insert were close to those for solid water. Baselines can be established using the average of the initial 5∼10 measurements.

Conclusions: Using longitudinal data, we estimated the action limits for CT number accuracy in the spectral images. This paves the way for establishing a comprehensive quality control program for spectral CT imaging.

Keywords: ACR phantom; dual-energy CT; quality control.