Outcome of stereo-electroencephalography with single-unit recording in drug-refractory epilepsy

J Neurosurg. 2023 May 26;139(6):1588-1597. doi: 10.3171/2023.4.JNS222633. Print 2023 Dec 1.

Abstract

Objective: The aim of this study was to evaluate the utility and safety of "hybrid" stereo-electroencephalography (SEEG) in guiding epilepsy surgery and in providing information at single-neuron levels (i.e., single-unit recording) to further the understanding of the mechanisms of epilepsy and the neurocognitive processes unique to humans.

Methods: The authors evaluated 218 consecutive patients undergoing SEEG procedures from 1993 through 2018 at a single academic medical center to assess the utility and safety of this technique in both guiding epilepsy surgery and providing single-unit recordings. The hybrid electrodes used in this study contained macrocontacts and microwires to simultaneously record intracranial EEG and single-unit activity (hybrid SEEG). The outcomes of SEEG-guided surgical interventions were examined, as well as the yield and scientific utility of single-unit recordings in 213 patients who participated in the research involving single-unit recordings.

Results: All patients underwent SEEG implantation by a single surgeon and subsequent video-EEG monitoring (mean of 10.2 electrodes per patient and 12.0 monitored days). Epilepsy networks were localized in 191 (87.6%) patients. Two clinically significant procedural complications (one hemorrhage and one infection) were noted. Of 130 patients who underwent subsequent focal epilepsy surgery with a minimum 12-month follow-up, 102 (78.5%) underwent resective surgery and 28 (21.5%) underwent closed-loop responsive neurostimulation (RNS) with or without resection. Seizure freedom was achieved in 65 (63.7%) patients in the resective group. In the RNS group, 21 (75.0%) patients achieved 50% or greater seizure reduction. When the initial period of 1993 through 2013 before responsive neurostimulator implantation in 2014 was compared with the subsequent period of 2014 through 2018, the proportion of SEEG patients undergoing focal epilepsy surgery grew from 57.9% to 79.7% due to the advent of RNS, despite a decline in focal resective surgery from 55.3% to 35.6%. A total of 18,680 microwires were implanted in 213 patients, resulting in numerous significant scientific findings. Recent recordings from 35 patients showed a yield of 1813 neurons, with a mean yield of 51.8 neurons per patient.

Conclusions: Hybrid SEEG enables safe and effective localization of epileptogenic zones to guide epilepsy surgery and provides unique scientific opportunities to investigate neurons from various brain regions in conscious patients. This technique will be increasingly utilized due to the advent of RNS and may prove a useful approach to probe neuronal networks in other brain disorders.

Keywords: Behnke-Fried electrodes; complications; depth electrodes; electroencephalography; epilepsy; human cognition; intracranial EEG; stereo-EEG.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, N.I.H., Extramural

MeSH terms

  • Drug Resistant Epilepsy* / surgery
  • Electrodes, Implanted
  • Electroencephalography / methods
  • Epilepsies, Partial* / surgery
  • Epilepsy* / surgery
  • Humans
  • Retrospective Studies
  • Seizures / surgery
  • Stereotaxic Techniques
  • Treatment Outcome