Background and aim: Measurement of cardiac index (CI) is crucial in the hemodynamic assessment of critically ill patients in the intensive care unit (ICU). The most reliable trans-thoracic echocardiography (TTE) technique for CI estimation is the left ventricular outflow tract (LVOT) Doppler method that requires, among other parameters, the LVOT cross-sectional area (CSA) measurement. However, inherent and practical disadvantages, mostly related to the ICU setting, hamper LVOT-CSA assessment. In this study, we aimed to validate a simplified formula, leveraging on LVOT-velocity time integral (VTI) and heart rate (HR) only, for non-invasive estimation of CI in ICU patients.
Methods and results: We prospectively enrolled 50 consecutive patients admitted to our ICU requiring pulmonary artery catheterization (PAC) over a one-year period. For each patient we measured the CI by PAC (CIPAC) and TTE. The latter was obtained both with the "traditional formula" (traditional CITTE), requiring LVOT-CSA assessment, and our new "simplified formula" (simplified CITTE). The correlation between the simplified CITTE and CIPAC was strong (r = 0.81) and resulted significantly greater than the traditional CITTE and CIPAC correlation (r = 0.70; p < 0.05 for Pearson r coefficients comparison). Both TTE-based CI showed an acceptable agreement (+0.19 ± 0.48 L/min/m2 for simplified CITTE and - 0.18 ± 0.58 L/min/m2 for traditional CITTE) with the reference CIPAC.
Conclusion: In this study, we validated a practical approach, leveraging on TTE LVOT-VTI and HR only, for non-invasive estimation of CI in ICU patients.
Keywords: Cardiac index; Echocardiography; Intensive care unit; Left ventricular outflow tract; Pulmonary artery catheterization; Simplified formula.
Copyright © 2022 Elsevier B.V. All rights reserved.