Background: The entrainment response, defined as the difference between the postpacing interval and the tachycardia cycle length (TCL) recorded from a mapping catheter, allows to track down the components of the tachycardia loop.
Objectives: The aim of this study was to evaluate if the postpacing interval measured simultaneously from multiple sites that are remote from the pacing site (PPIR) could be clinically useful in mapping re-entrant circuits.
Methods: Ninety-two episodes of entrainment response in 29 patients with different macro-re-entrant tachycardias were evaluated using a standardized entrainment protocol. The spatial distribution of different values of PPIR-TCL in a simulation and a computational model of an entrained re-entrant tachycardia was also analyzed.
Results: The PPIR exceeded TCL by more than 20 milliseconds only if both pacing and recording sites were outside the tachycardia circuit. The PPIR-TCL at in-circuit sites was always ≤20 milliseconds. Sites with negative PPIR-TCL values were found either outside or inside the tachycardia circuit.
Conclusions: Assessment of entrainment response from catheters remote from the pacing site may enhance spatial mapping of the tachycardia circuit. The PPIR-TCL above 20 milliseconds has an excellent positive predictive value in identifying sites outside the tachycardia circuit.
Keywords: electrophysiology mapping; entrainment; re-entry.
Copyright © 2022 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.