Background: Aortic valve replacement (AVR) is one of the most common open-heart surgical procedures. The durability of the tissue valve in the aortic position is crucial in AVR and transcatheter AVR. We reviewed structural valve deterioration using echocardiographic follow-up in three types of surgical aortic tissue valves.
Methods: A retrospective analysis was conducted where hemodynamic deterioration was evaluated and compared using transthoracic echocardiography, including pressure gradients and effective orifice area. Kaplan-Meier analyses were used to summarize the time to failure.
Results: The study included 133 Trifecta, 156 Epic, and 321 Magna Ease valves. Seventy-six percent (1941/2551) of patients had to be excluded due to insufficient echo data. Through univariate analysis, 34% (216/610) of valves met deterioration criteria after 24 months. Unadjusted survival curves showed a significant difference between valves (p ≤ .001), with a longer mean time to deterioration for the Magna Ease versus Trifecta and Epic of 68.9 versus 50.1 and 38.2 months, respectively. A Cox proportional hazard analysis found worse hazard ratios of 1.69 (p ≤ .04) and 2.4 (p ≤ .01) for Trifecta versus Magna and Epic versus Trifecta, respectively.
Conclusion: All three valve types demonstrated structural valve deterioration on echocardiographic follow-up with significant differences in rate. The Magna Ease appeared to have the highest durability, and the Epic the lowest. Further investigation is warranted to confirm the results in a larger multicenter study.
Keywords: aorta and great vessels; valve repair/replacement.
© 2022 Wiley Periodicals LLC.