Objective: To determine if laser speckle contrast imaging (LSCI) mitigates variations and subjectivity in the use and interpretation of indocyanine green (ICG) fluorescence in the current visualization paradigm of real-time intraoperative tissue blood flow/perfusion in clinically relevant scenarios.
Methods: De novo laparoscopic imaging form-factor detecting real-time blood flow using LSCI and blood volume by near-infrared fluorescence (NIRF) of ICG was compared to ICG NIRF alone, for dye-less real-time visualization of tissue blood flow/perfusion. Experienced surgeons examined LSCI and ICG in segmentally devascularized intestine, partial gastrectomy, and the renal hilum across six porcine models. Precision and accuracy of identifying demarcating lines of ischemia/perfusion in tissues were determined in blinded subjects with varying levels of surgical experience.
Results: Unlike ICG, LSCI perfusion detection was real time (latency < 150 ms: p < 0.01), repeatable and on-demand without fluorophore injection. Operating surgeons (n = 6) precisely and accurately identified concordant demarcating lines in white light, LSCI, and ICG modes immediately. Blinded subjects (n = 21) demonstrated similar spatial-temporal precision and accuracy with all three modes ≤ 2 min after ICG injection, and discordance in ICG mode at ≥ 5 min in devascularized small intestine (p < 0.0001) and in partial gastrectomy (p < 0.0001).
Conclusions: Combining LSCI for near real-time blood flow detection with ICG fluorescence for blood volume detection significantly improves precision and accuracy of perfusion detection in tissue locations over time, in real time, and repeatably on-demand than ICG alone.
Keywords: Anastomosis; Dye-less real-time perfusion; Indocyanine green; Laser speckle contrast imaging; Near-infrared fluorescence.
© 2022. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.