Background/purpose: In daily plan adaptation the radiotherapy treatment plan is adjusted just prior to delivery. A simple approach is taking the planning objectives of the reference plan and directly applying these in re-optimization. Here we present a tested method to verify whether daily adaptation without tweaking of the objectives can maintain the plan quality throughout treatment.
Materials/methods: For fifteen rectal cancer patients, automated treatment planning was used to generate plans mimicking manual reference plans on the planning scans. For 74 fraction scans (4-5 per patient) an automated plan and a daily adapted plan were generated, where the latter re-optimizes the reference plan objectives without any tweaking. To evaluate the robustness of the daily adaptation, the adapted plans were compared to the autoplanning plans.
Results: Median differences between the autoplanning plans on the planning scans and the reference plans were between -1 and 0.2 Gy. The largest interquartile range (1 Gy) was seen for the Lumbar Skin D2%. For the daily scans the PTV D2% and D98% differences between autoplanning and adapted plans were within 0.7 Gy, with mean differences within 0.3 Gy. Positive differences indicate higher values were obtained using autoplanning. For the Bowelarea + Bladder and the Lumbar Skin the D2% and Dmean differences were all within 2.6 Gy, with mean differences between -0.9 and 0.1 Gy.
Conclusion: Automated treatment planning can be used to benchmark daily adaptation techniques. The investigated adaptation workflow can robustly perform high quality adaptations without daily adjusting of the patient-specific planning objectives for rectal cancer radiotherapy.
Keywords: Automated treatment planning; MRI guided radiotherapy; Online plan adaptation.