Repeatability of QuantiFERON-TB gold plus testing utilizing microparticle chemiluminescence

J Immunol Methods. 2022 Oct:509:113340. doi: 10.1016/j.jim.2022.113340. Epub 2022 Aug 27.

Abstract

Background: Detection of latent Mycobacterium tuberculosis (LTBI) in patients is important to prevent active infection and the spread of disease, particularly in vulnerable patient populations. In 2020, a kit on the high throughput Liaison XL (DiaSorin) became commercially available for the analysis of QuantiFERON-TB Gold Plus assay (Qiagen). Pilot testing indicated suboptimal repeatability of some samples with this assay. This study provides an extensive assessment of repeatability with DiaSorin system.

Results: Repeat testing of 481 IGRA positive samples, demonstrated substantial variability upon repeat analysis. Repeat results for TB1 and TB2 tubes, showed 73.73% and 72.82% concordance with initial results, respectively. TB1 and TB2 tube values minus the nil (IU/mL) were significantly higher in samples that were repeat positive (p < 0.001). Repeat results had better concordance with initial results if both TB1 and TB2 tubes were positive. Samples with TB1 tube values minus the nil (IU/mL) ≥ 4.54 and TB2 tube values minus the nil (IU/mL) ≥ 4.78 were found to always repeat positive. Assigning a threshold of 1.55 IU/mL for the TB1 tube value minus the nil and 1.45 IU/mL for the TB2 tube value minus the nil yielded a positive predictive value ≥95%.

Conclusion: These results identified a potential role for retesting of select IGRA positive samples on the Diasorin Liaison XL platform due to the high proportion of samples that show a lack of repeatability. Additionally, we identified a threshold that would determine samples most likely to repeat test positive and which samples should be retested.

Keywords: IGRA; LTBI; TB; Tuberculosis.

MeSH terms

  • Humans
  • Interferon-gamma Release Tests / methods
  • Latent Tuberculosis* / diagnosis
  • Latent Tuberculosis* / microbiology
  • Luminescence
  • Mycobacterium tuberculosis*
  • Predictive Value of Tests