Advanced Online Monitoring of In Vitro Human 3D Full-Thickness Skin Equivalents

Pharmaceutics. 2022 Jul 8;14(7):1436. doi: 10.3390/pharmaceutics14071436.

Abstract

Skin equivalents and skin explants are widely used for dermal penetration studies in the pharmacological development of drugs. Environmental parameters, such as the incubation and culture conditions affect cellular responses and thus the relevance of the experimental outcome. However, available systems such as the Franz diffusion chamber, only measure in the receiving culture medium, rather than assessing the actual conditions for cells in the tissue. We developed a sampling design that combines open flow microperfusion (OFM) sampling technology for continuous concentration measurements directly in the tissue with microfluidic biosensors for online monitoring of culture parameters. We tested our design with real-time measurements of oxygen, glucose, lactate, and pH in full-thickness skin equivalent and skin explants. Furthermore, we compared dermal penetration for acyclovir, lidocaine, and diclofenac in skin equivalents and skin explants. We observed differences in oxygen, glucose, and drug concentrations in skin equivalents compared to the respective culture medium and to skin explants.

Keywords: full-thickness skin equivalent; glucose; lactate; open flow microperfusion; oxygen; pH; penetration tests; pharmacokinetics; sensors; skin explants.