Purpose: Quantitative multi-parameter mapping (MPM) has been shown to provide good longitudinal and cross-sectional reproducibility for clinical research. Unfortunately, acquisition times (TAs) are typically infeasible for routine scanning at high resolutions.
Methods: A fast whole-brain MPM protocol based on interleaved multi-shot 3D-EPI with controlled aliasing (SC-EPI) at 3T and 7T is proposed and compared with MPM using a standard spoiled gradient echo (FLASH) sequence. Four parameters (R1 , PD, , and MTsat) were measured in less than 3 min at 1 mm isotropic resolution. Five subjects went through the same scanning sessions twice at each scanner. The intra-subject coefficient of variation (scan-rescan) (CoV) was estimated for each protocol and scanner to assess the longitudinal reproducibility.
Results: At 3T, the CoV of SC-EPI ranged between 1.2%-4.8% for PD and R1 , 2.8%-10.6% for and MTsat, which was comparable with FLASH (0.6%-4.9% for PD and R1 , 2.6%-11.3% for and MTsat). At 7T, where the SC-EPI TA was reduced to ∼2 min, the CoV of SC-EPI (1.4%-10.6% for PD, R1 , and ) was 1.2-2.4 times larger than the CoV of FLASH (1.0%-15%) and MTsat showed much higher variability across subjects. The SC-EPI-MPM protocol at 3T showed high reproducibility and yielded stable quantitative maps at a clinically feasible resolution and scan time, whereas at 7T, MT saturation homogeneity needs to be improved.
Conclusion: SC-EPI-based MPM is feasible as an additional MRI modality in clinical or population studies where the parameters offer great potential as biomarkers.
Keywords: EPI; MPM; MT; PD; T1; T2*.
© 2022 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.