Radiolucent zones of biodegradable magnesium-based screws in children and adolescents-a radiographic analysis

Arch Orthop Trauma Surg. 2023 May;143(5):2297-2305. doi: 10.1007/s00402-022-04418-0. Epub 2022 Mar 29.

Abstract

Introduction: Albeit the implantation of magnesium-based biodegradable implants can avoid a second surgery for implant removal, the postoperative occurrence of radiolucent zones around these implants based on corrosion processes has not been previously investigated in children and adolescents. We sought to characterize the distinct temporal and spatial dynamics for magnesium-based implants based on standard clinical routine radiographs.

Materials and methods: We retrospectively analyzed 29 patients, treated with magnesium-based compression screws (MAGNEZIX® CS 2.7 mm, CS 3.2 mm, CSC 4.8 mm; Syntellix AG) for fracture fixation, osteotomy, or osteochondral refixation. During a follow-up examination, the clinical and functional status was evaluated. Based on digital radiographs, the ratio of the area of the radiolucent zone and that of the screw was evaluated to assess implant degradation at two follow-up visits (i.e., after 6-8 weeks and 12-24 weeks).

Results: In 29 patients (16/29 females, 14.03 ± 2.13 years), a total of 57 implants were evaluated that were used for osteotomy (n = 13, screws n = 26), fracture fixation (n = 9, screws n = 18), or osteochondral refixation (n = 7, screws n = 13). All patients healed without complications and regained full function. Radiolucent zones were observed in 27/29 patients at the first follow-up, with significantly decreased ratios at the second follow-up (2.10 ± 0.55 vs 1.64 ± 0.60, p = 0.0006). Regression analyses were performed to assess the temporal dynamics of radiolucent zones and revealed significant logarithmic developments for the 2.7 mm and 3.2 mm screws, marked by a strong ratio decrease during the first weeks and an almost complete disappearance after approximately 100 days and 200 days, respectively. In contrast, the ceramic-coated 4.8 mm screws presented a significant linear and slower decrease of radiolucent zones.

Conclusion: Radiolucent zones are a common phenomenon in the course of implant degradation. However, they represent a self-limiting phenomenon. Remarkably, neither implant failure nor affected implant function is noted in this context. Yet, the microstructural changes accompanying the presence of radiolucent zones remain to be analyzed by three-dimensional high-resolution imaging.

Keywords: Children; Magnesium screws; Osteochondral refixation; Osteosynthesis; Osteotomy; Radiolucent zones.

MeSH terms

  • Absorbable Implants
  • Adolescent
  • Bone Screws* / adverse effects
  • Child
  • Female
  • Humans
  • Magnesium* / adverse effects
  • Osteotomy / methods
  • Radiography
  • Retrospective Studies

Substances

  • Magnesium