Using an Online Disturbance Rejection and Anticipation System to Reduce Hyperglycemia in a Fully Closed-Loop Artificial Pancreas System

J Diabetes Sci Technol. 2022 Jan;16(1):52-60. doi: 10.1177/19322968211059159. Epub 2021 Dec 3.

Abstract

Introduction: Hyperglycemia following meals is a recurring challenge for people with type 1 diabetes, and even the most advanced available automated systems currently require manual input of carbohydrate amounts. To progress toward fully automated systems, we present a novel control system that can automatically deliver priming boluses and/or anticipate eating behaviors to improve postprandial full closed-loop control.

Methods: A model predictive control (MPC) system was enhanced by an automated bolus system reacting to early glucose rise and/or a multistage MPC (MS-MPC) framework to anticipate historical patterns. Priming was achieved by detecting large glycemic disturbances, such as meals, and delivering a fraction of the patient's total daily insulin (TDI) modulated by the disturbance's likelihood (bolus priming system [BPS]). In the anticipatory module, glycemic disturbance profiles were generated from historical data using clustering to group days with similar behaviors; the probability of each cluster is then evaluated at every controller step and informs the MS-MPC framework to anticipate each profile. We tested four configurations: MPC, MPC + BPS, MS-MPC, and MS-MPC + BPS in simulation to contrast the effect of each controller module.

Results: Postprandial time in range was highest for MS-MPC + BPS: 60.73 ± 25.39%, but improved with each module: MPC + BPS: 56.95±25.83 and MS-MPC: 54.83 ± 26.00%, compared with MPC: 51.79 ± 26.12%. Exposure to hypoglycemia was maintained for all controllers (time below 70 mg/dL <0.5%), and improvement came primarily from a reduction in postprandial time above range (MS-MPC + BPS: 39.10 ± 25.32%, MPC + BPS: 42.99 ± 25.81%, MS-MPC: 45.09 ± 25.96%, MPC: 48.18 ± 26.09%).

Conclusions: The BPS and anticipatory disturbance profiles improved blood glucose control and were most efficient when combined.

Keywords: artificial pancreas; automated insulin delivery; behavioral patterns; disturbance mitigation; meal detection.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Blood Glucose
  • Diabetes Mellitus, Type 1*
  • Humans
  • Hyperglycemia* / prevention & control
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Insulin Infusion Systems
  • Pancreas, Artificial*

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin