Anterior segment biometric measurements explain misclassifications by a deep learning classifier for detecting gonioscopic angle closure

Br J Ophthalmol. 2023 Mar;107(3):349-354. doi: 10.1136/bjophthalmol-2021-319058. Epub 2021 Oct 6.

Abstract

Background/aims: To identify biometric parameters that explain misclassifications by a deep learning classifier for detecting gonioscopic angle closure in anterior segment optical coherence tomography (AS-OCT) images.

Methods: Chinese American Eye Study (CHES) participants underwent gonioscopy and AS-OCT of each angle quadrant. A subset of CHES AS-OCT images were analysed using a deep learning classifier to detect positive angle closure based on manual gonioscopy by a reference human examiner. Parameter measurements were compared between four prediction classes: true positives (TPs), true negatives (TNs), false positives (FPs) and false negatives (FN). Logistic regression models were developed to differentiate between true and false predictions. Performance was assessed using area under the receiver operating curve (AUC) and classifier accuracy metrics.

Results: 584 images from 127 participants were analysed, yielding 271 TPs, 224 TNs, 77 FPs and 12 FNs. Parameter measurements differed (p<0.001) between prediction classes among anterior segment parameters, including iris curvature (IC) and lens vault (LV), and angle parameters, including angle opening distance (AOD). FP resembled TP more than FN and TN in terms of anterior segment parameters (steeper IC and higher LV), but resembled TN more than TP and FN in terms of angle parameters (wider AOD). Models for detecting FP (AUC=0.752) and FN (AUC=0.838) improved classifier accuracy from 84.8% to 89.0%.

Conclusions: Misclassifications by an OCT-based deep learning classifier for detecting gonioscopic angle closure are explained by disagreement between anterior segment and angle parameters. This finding could be used to improve classifier performance and highlights differences between gonioscopic and AS-OCT definitions of angle closure.

Keywords: angle; diagnostic tests/investigation; glaucoma; imaging.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anterior Eye Segment / diagnostic imaging
  • Biometry
  • Deep Learning*
  • Glaucoma, Angle-Closure* / diagnosis
  • Gonioscopy
  • Humans
  • Intraocular Pressure
  • Prospective Studies
  • Tomography, Optical Coherence / methods

Substances

  • 2-(N-cyclohexylamino)ethanesulfonic acid