Background: In endovascular aneuysm repair (EVAR) procedures, medical instruments are currently navigated with a two-dimensional imaging based guidance requiring X-rays and contrast agent.
Methods: Novel approaches for obtaining the three-dimensional instrument positions are introduced. Firstly, a method based on fibre optical shape sensing, one electromagnetic sensor and a preoperative computed tomography (CT) scan is described. Secondly, an approach based on image processing using one 2D fluoroscopic image and a preoperative CT scan is introduced.
Results: For the tracking based method, average errors from 1.81 to 3.13 mm and maximum errors from 3.21 to 5.46 mm were measured. For the image-based approach, average errors from 3.07 to 6.02 mm and maximum errors from 8.05 to 15.75 mm were measured.
Conclusion: The tracking based method is promising for usage in EVAR procedures. For the image-based approach are applications in smaller vessels more suitable, since its errors increase with the vessel diameter.
Keywords: 2D/3D registration; 3D localisation; computer-assisted surgery; electromagnetic tracking system; endovascular procedures; fibre optical shape sensing.