Accuracy of matrix-assisted laser desorption ionization time-of-flight mass spectrometry for direct bacterial identification from culture-positive urine samples

Ann Transl Med. 2021 Apr;9(8):647. doi: 10.21037/atm-20-7310.

Abstract

Background: Urinary tract infection (UTI) is one of the most frequent reasons for antimicrobial therapy. In typical clinical setting, 18-48 h is needed to identify pathogens by urine culture. A rapid method for pathogenic UTI diagnosis by matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been developed in recent years.

Methods: This meta-analysis systematically evaluated the accuracy of MALDI-TOF MS for direct identification of bacteria from culture-positive urine samples. We queried the electronic database of Medline and Web of Science to obtain relevant articles.

Results: Nineteen articles involving 4,579 isolates were included after final selection in the meta-analysis. The random-effects pooled identification accuracy of MALDI-TOF MS was 0.82 with 95% confidence interval of 0.79 to 0.86 at the species level. For Gram-negative isolates, the correct identification performance of the species ranged from 0.54 to 0.98, with a cumulative rate of 0.87 (95% CI: 0.83 to 0.91). For Gram-positive isolates, the correct identification rate ranged from 0.32 to 0.80, with a cumulative rate of 0.59 (95% CI: 0.49 to 0.68).

Conclusions: MALDI-TOF MS provides a reliable direct identification of bacteria, particularly in cases of Gram-negative isolates, from clinical urine specimens. Nevertheless, the identification accuracy of this method is moderate for Gram-positive bacteria.

Keywords: Accuracy; culture-positive urine samples; direct detection; matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS).