Safety and Efficacy of Paclitaxel-Eluting Balloon Angioplasty for Dysfunctional Hemodialysis Access: A randomized trial Comparing with Angioplasty Alone

J Vasc Interv Radiol. 2021 Mar;32(3):350-359.e2. doi: 10.1016/j.jvir.2020.10.030. Epub 2021 Jan 19.

Abstract

Purpose: To assess whether angioplasty of hemodialysis access (HA) stenosis with a drug-coated balloon (DCB) would prevent restenosis in comparison with plain-balloon percutaneous transluminal angioplasty (PTA).

Materials and methods: This prospective randomized clinical trial enrolled 120 patients with dysfunctional arteriovenous fistulae (n = 109) and grafts (n = 11), due to a ≥50% stenosis between March 2014 and April 2018. All patients underwent high-pressure balloon angioplasty and were then randomized to either DCB (n = 60) or PTA (n = 60). Patients were followed-up for 1 year, and angiography was performed 6 months after angioplasty. The primary endpoint was the late lumen loss (LLL) at 6 months. Secondary endpoints included other angiographic parameters at 6 months and HA failures, adverse event, and mortality at 12 months. Continuous variables were compared with a Student t-test, and Kaplan-Meier curves were used for freedom from HA failure and for mortality.

Results: LLL in the DCB and in the PTA group were 0.64 mm ± 1.20 and 1.13 mm ± 1.51, respectively (P = .082, adjusted P = .0498). DCB was associated with lower percentage stenosis (54.2% ± 19.3 vs 61.7% ± 18.2; P = .047) and binary restenosis ≥50% (56.5% vs 81.1%; P = .009) than PTA. The number of HA failures after 12 months was lower for DCB than for PTA (45% vs 66.7%; P = .017). Mortality at 12 months was 10% and 8.3% in the DCB and PTA groups, respectively (P = .75).

Conclusions: Despite LLL improvement that failed to reach statistical significance, this study demonstrated decreased incidence and severity of restenosis with DCB compared with PTA to treat dysfunctional HA.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Angioplasty, Balloon / adverse effects
  • Angioplasty, Balloon / instrumentation*
  • Angioplasty, Balloon / mortality
  • Arteriovenous Shunt, Surgical / adverse effects*
  • Arteriovenous Shunt, Surgical / mortality
  • Blood Vessel Prosthesis Implantation / adverse effects*
  • Blood Vessel Prosthesis Implantation / mortality
  • Cardiovascular Agents / administration & dosage*
  • Coated Materials, Biocompatible*
  • Female
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / mortality
  • Graft Occlusion, Vascular / physiopathology
  • Graft Occlusion, Vascular / therapy*
  • Humans
  • Male
  • Middle Aged
  • Paclitaxel / administration & dosage*
  • Prospective Studies
  • Quebec
  • Recurrence
  • Renal Dialysis*
  • Risk Factors
  • Single-Blind Method
  • Time Factors
  • Treatment Outcome
  • Vascular Access Devices*
  • Vascular Patency

Substances

  • Cardiovascular Agents
  • Coated Materials, Biocompatible
  • Paclitaxel