Dosimetric response of a glass dosimeter in proton beams: LET-dependence and correction factor

Phys Med. 2021 Jan:81:147-154. doi: 10.1016/j.ejmp.2020.12.001. Epub 2021 Jan 15.

Abstract

A radiophotoluminescent glass dosimeter (RGD) is widely used in postal audit system for photon beams in Japan. However, proton dosimetry in RGDs is scarcely used owing to a lack of clarity in their response to beam quality. In this study, we investigated RGD response to beam quality for establishing a suitable linear energy transfer (LET)-corrected dosimetry protocol in a therapeutic proton beam. The RGD response was compared with ionization chamber measurement for a 100-225 MeV passive proton beam. LET of the measurement points was calculated by the Monte Carlo method. An LET-correction factor, defined as a ratio between the non-corrected RGD dose and ionization chamber dose, of 1.226×(LET)-0.171 was derived for the RGD response. The magnitude of the LET-dependence of RGD increased with LET; for an LET of 8.2 keV/μm, the RGD under-response was up to 16%. The coefficient of determination, mean difference ± SD of non-corrected RGD dose, residual range-corrected RGD dose, and LET-corrected RGD dose to the ionization chamber are 0.923, 3.7 ± 4.2%, -2.4 ± 7.5%, and 0.04 ± 2.1%, respectively. The LET-corrected RGD dose was within 5% of the corresponding ionization chamber dose at all energies until 200 MeV, where it was 5.3% lower than the ionization chamber dose. A corrected LET-dependence of RGD using a correction factor based on a power function of LET and precise dosimetric verification close to the maximum LET were realized here. We further confirmed establishment of an accurate postal audit under various irradiation conditions.

Keywords: Linear energy transfer; Postal audit; Proton beam; Radiophotoluminescent glass dosimeter; Under-response correction.

MeSH terms

  • Japan
  • Linear Energy Transfer*
  • Monte Carlo Method
  • Proton Therapy*
  • Protons
  • Radiation Dosimeters
  • Radiometry

Substances

  • Protons