Considerations in the Use of Gravitational Valves in the Management of Hydrocephalus. Some Lessons Learned with the Dual-Switch Valve

J Clin Med. 2021 Jan 12;10(2):246. doi: 10.3390/jcm10020246.

Abstract

In the past decade, there has been a clear trend towards better outcomes in patients with hydrocephalus, especially those with normal pressure hydrocephalus (NPH). This is partly due to the availability of more sophisticated hardware and a better understanding of implants. However, there is little evidence to show the superiority of a specific type of valve over another. The most commonly reported consequence of hydrodynamic mismatch is shunt over-drainage. Simple differential pressure valves, with a fixed opening pressure or even adjustable valves, lead to non-physiologic intraventricular pressure (IVP) as soon as the patient moves into an upright posture. These valves fail to maintain IVP within physiological limits due to the changes in hydrostatic pressure in the drainage system. To solve this problem more complex third-generation hydrostatic valves have been designed. These gravitational devices aim to reduce flow through a shunt system when the patient is upright but there are important technical differences between them. Here we review the main characteristics of the Miethke® Dual-Switch valve, which includes two valve chambers arranged in parallel: a low-opening pressure valve, designed for working in the supine position, and a second high-opening pressure valve, which starts working when the patient assumes the upright position. This paper specifies the main advantages and drawbacks of this device and provide a series of recommendations for its use. The discussion of this specific gravitational valve allows us to emphasize the importance of using gravitational control in implanted shunts and some the caveats neurosurgeons should take into consideration when using gravitational devices in patients with hydrocephalus. The correct function of any gravitational device depends on adequate device implantation along the vertical body axis. Misalignment from the vertical axis equal to or more than 45° might eliminate the beneficial effect of these devices.

Keywords: Dual-Switch valve; complications; gravitational valves; hydrocephalus; hydrostatic valves; shunt dysfunction.

Publication types

  • Review