Breath-hold and free-breathing quantitative assessment of biventricular volume and function using compressed SENSE: a clinical validation in children and young adults

J Cardiovasc Magn Reson. 2020 Jul 27;22(1):54. doi: 10.1186/s12968-020-00642-y.

Abstract

Background: Although the breath-hold cine balanced steady state free precession (bSSFP) imaging is well established for assessment of biventricular volumes and function, shorter breath-hold times or no breath-holds are beneficial in children and severely ill or sedated patients.

Methods: Clinical cardiovascular magnetic resonance (CMR) examinations from September 2019 to October 2019 that included breath-hold (BH) and free-breathing (FB) cine bSSFP imaging accelerated using compressed sensitivity encoding (C-SENSE) factor of 3 in addition to the clinical standard BH cine bSSFP imaging using SENSE factor of 2 were analyzed retrospectively. Patients with structurally normal hearts who could perform consistent BHs were included. Aortic flow measured by phase contrast acquisition was used as a reference for the left ventricular (LV) stroke volume. Comparative analysis was performed for evaluation of biventricular volumes and function, imaging times, quantitative image quality, and qualitative image scoring.

Results: There were 26 patients who underwent all three cine scans during the study period (16.7 ± 6.4 years, body surface area (BSA) 1.6 ± 0.4 m2, heart rate 83 ± 7 beats/min). BH durations of 8 ± 1 s with C-SENSE = 3 were significantly shorter (p < 0.001) by 33% compared to 12 ± 1 s with SENSE = 2. Actual scan time for BH SENSE (4.9 ± 1.2 min) was comparable to that with FB C-SENSE (5.2 ± 1.5 min; p= NS). Biventricular stroke volume and ejection fraction, and LV mass computed using all three sequences were comparable. There was a small but statistically significant (p < 0.05) difference in LV end-diastolic volume (- 3.0 ± 6.8 ml) between BH SENSE and FB C-SENSE. There was a small but statistically significant (p < 0.005) difference in end-diastolic LV (- 5.0 ± 7.7 ml) and RV (- 6.0 ± 8.5 ml) volume and end-systolic LV (- 3.2 ± 4.3 ml) and RV(- 4.2 ± 6.8 ml) volumes between BH C-SENSE and FB C-SENSE. The LV stroke volumes from all three sequences had excellent correlations (r = 0.96, slope = 0.98-1.02) with aortic flow, with overestimation by 2.7 (5%) to 4.6 (8%) ml/beat. The image quality score was Excellent (16 of 26) to Good (10 of 26) with BH SENSE, Excellent (13 of 26) to Good (13 of 26) with BH C-SENSE, and Excellent (3 of 26) to Good (21 of 26) to Adequate (2 of 26) with FB C-SENSE.

Conclusions: Image quality and ventricular volumetric and functional indices using either BH or FB C-SENSE cine bSSFP imaging were comparable to standard BH SENSE cine bSSFP imaging while maintaining nominally identical spatio-temporal resolution. This accelerated image acquisition provides an alternative to accommodate patients with impaired BH capacity.

Keywords: Children; Compressed SENSE; Free-breathing cine; Left ventricular indices; Pediatric; Right ventricular indices.

Publication types

  • Comparative Study
  • Validation Study
  • Video-Audio Media

MeSH terms

  • Adolescent
  • Adult
  • Age Factors
  • Breath Holding*
  • Cardiac-Gated Imaging Techniques
  • Child
  • Female
  • Heart Diseases / diagnostic imaging*
  • Heart Diseases / etiology
  • Heart Diseases / physiopathology
  • Heart Ventricles / diagnostic imaging*
  • Heart Ventricles / physiopathology
  • Humans
  • Image Interpretation, Computer-Assisted*
  • Magnetic Resonance Imaging, Cine*
  • Male
  • Predictive Value of Tests
  • Reproducibility of Results
  • Retrospective Studies
  • Stroke Volume*
  • Time Factors
  • Ventricular Function, Left*
  • Ventricular Function, Right*
  • Young Adult