Introduction: In cardiac resynchronization therapy, pacing the left ventricle (LV) at sites of prolonged electrical delay is associated with better outcomes. We sought to characterize the interrelationships between intrinsic, right-ventricular (RV)-paced, and LV-paced interventricular delays.
Methods and results: The following electrical timings were measured at implantation for all electrodes of the LV quadripolar leads: QLV, interventricular delay in intrinsic rhythm (RVs-LVs), in RV-paced rhythm (RVp-LVs), and in LV-paced rhythm (LVp-RVs). We included 32 patients (78% men, age 72 years, LV ejection fraction 29%, left bundle branch block 84%). QLV and RVs-LVs were correlated (R2 = .72, p < .0001), as were RVs-LVs and RVp-LVs (R2 = .27, p = .002) and RVp-LVs and LVp-RVs (R2 = .60, p < .001). Direction of activation along the four LV lead electrodes was concordant between RVs-LVs and RVp-LVs in only 17 (53%) patients. The latest-activated electrodes in RVs-LVs and RVp-LVs were concordant in 26 (81%) patients, adjacent in 3 (9%) patients, and remote in 3 (9%) patients. Biventricular-paced QRS duration varied by more than 10 ms between the two electrodes in half of the patients with dissimilar latest electrodes. Among the seven echocardiographic nonresponders at 6 months, the programmed electrode was remote from the latest electrode in RVs-LVs in five patients and in RVp-LVs in three patients.
Conclusion: Intrinsic and RV-paced interventricular electrical delays are correlated, but there is substantial heterogeneity between patients. The latest-activated electrode may be different between RVs-LVs and RVp-LVs, and this might have important implications in selecting the optimal LV vector.
Keywords: activation propagation; cardiac resynchronization therapy; correlation; interventricular delays; latest-activated electrode; local electrical delay.
© 2020 Wiley Periodicals LLC.