Background: Tissue growth over covered branches is a leading cause of delayed thrombotic complications after flow-diverter stenting (FDS). Due to insufficient resolution, no imaging modality is clinically available to monitor this phenomenon.
Objective: To evaluate high-frequency optical coherence tomography (HF-OCT), a novel intravascular imaging modality designed for the cerebrovascular anatomy with a resolution approaching 10 microns, to monitor tissue growth over FDS in an arterial bifurcation model.
Methods: FDS were deployed in a rabbit model (n = 6), covering the aortic bifurcation. The animals were divided in different groups, receiving dual antiplatelet therapy (DAPT) (n = 4), aspirin only (n = 1), and no treatment (n = 1). HF-OCT data were obtained in vivo at 3 different time points in each animal. For each cross-sectional image, metal and tissue coverage of the jailed ostium was quantified. Scanning electron microscopy images of harvested arteries were subsequently obtained.
Results: Good quality HF-OCT data sets were successfully acquired at implant and follow-up. A median value of 41 (range 21-55) cross-sectional images were analyzed per ostium for each time point. Between 0 and 30 d after implant, HF-OCT analysis showed a significantly higher ostium coverage when DAPT was not given. After 30 d, similar growth rates were found in the DAPT and in the aspirin group. At 60 d, a coverage of 90% was reached in all groups.
Conclusion: HF-OCT enables an accurate visualization of tissue growth over time on FDS struts. The use of FDS in bifurcation locations may induce a drastic reduction of the jailed-branch ostium area.
Keywords: Endothelialization; Flow diverter; Intracranial aneurysm; Optical coherence tomography.
Copyright © 2020 by the Congress of Neurological Surgeons.