Transcatheter closure of a perimembranous ventricular septal defect with Nit-Occlud Lê VSD Coil: A French multicentre study

Arch Cardiovasc Dis. 2020 Feb;113(2):104-112. doi: 10.1016/j.acvd.2019.11.004. Epub 2020 Jan 22.

Abstract

Background: Transcatheter perimembranous ventricular septal defect (pmVSD) closure remains challenging and is seldom used in France given the risk of atrioventricular block (AVB). pmVSD closure with the Nit-Occlud Lê VSD coil was recently introduced in France as an alternative to occluder devices.

Aims: To study the safety and feasibility of pmVSD closure with the Nit-Occlud Lê VSD coil.

Methods: All consecutives cases of pmVSD closure with the Nit-Occlud Lê VSD coil in 20 tertiary French centres were included between January 2015 and December 2018.

Results: Among 46 procedures in five centres, indications for pmVSD closure were left ventricle overload (76.1%), exertional dyspnoea (17.4%), history of infective endocarditis (4.3%) and mild pulmonary hypertension (2.2%). The median (interquartile [IQR]) age of the patients was 13.9 (5.7-31.8) years. Aneurismal tissue was identified in 91.3% of patients. VSD median (IQR) size was 8 (7-10) mm on the left ventricle side and 5 (4-6) mm on the right ventricle side. Implantation was successful in 40 patients (87.0%; 95% confidence interval [CI] 73.7-95.1%). Severe complications occurred in six patients (13.0%, 95% CI 4.9-26.3%), mainly severe haemolysis (8.7%, 95% CI 2.4-20.8%). One aortic valve lesion required surgical aortic valvuloplasty. Occurrence of severe complications was significantly related to the presence of haemolysis (P=0.001), residual shunt (P=0.007) and multi-exit VSD (P=0.005). Residual shunt was observed in 40% of cases with the implanted device shortly after closure and 15% after a median follow-up of 27 months. No immediate or delayed device embolization or complete AVB was recorded.

Conclusion: pmVSD closure with the Nit-Occlud Lê VSD Coil is feasible in older children and adults. However, residual shunting (leading to haemolysis) is a dreaded complication that should not be tolerated. pmVSD closure with the Nit-Occlud Lê VSD as a therapeutic strategy remains controversial and is limited to selected patients.

Keywords: Catheterization; Cathétérisme; Communication interventriculaire; Dilatation ventriculaire gauche; Fermeture de shunt; Left ventricle overload; Nit-Occlud Lê VSD; Shunt closure; Ventricular septal defect.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Adolescent
  • Adult
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Child
  • Child, Preschool
  • Feasibility Studies
  • Female
  • France
  • Heart Septal Defects, Ventricular / diagnostic imaging
  • Heart Septal Defects, Ventricular / physiopathology
  • Heart Septal Defects, Ventricular / therapy*
  • Hemodynamics
  • Hemolysis
  • Humans
  • Male
  • Patient Safety
  • Prosthesis Design
  • Recovery of Function
  • Risk Assessment
  • Risk Factors
  • Septal Occluder Device*
  • Time Factors
  • Treatment Outcome