Purpose: To assess the feasibility and reproducibility of T2 relaxation time measurements of the trapeziometacarpal joint (TM) and triangular fibrocartilage complex (TFCC) on healthy subjects at 1.5 T MR.
Materials and methods: Thirty-four healthy volunteers underwent an axial oblique multislice multiecho spin-echo sequence of the wrist at 1.5 T, with 10 of them having performed another MR scan on a different 1.5 T scanner. Regions of interest were independently manually drawn by two musculoskeletal radiologists to include the cartilaginous part of the TM and TFCC. Intra-observer, inter-observer and inter-scanner reproducibility of T2 relaxation time measurements was tested using the Bland-Altman method.
Results: The mean T2 values obtained by the two radiologists were 29.9 ± 6.5 ms and 30.0 ± 6.1 ms in the TM and 24.5 ± 2.3 ms and 24.6 ± 2.8 ms in the TFCC, respectively. The mean values of the second series of T2 measurements obtained by the senior radiologist were 29.9 ± 6.5 ms and 30.0 ± 6.3 ms in the TM and 24.3 ± 2.9 ms in the TFCC. Inter-observer reproducibility in the TM and in the TFCC was 76% and 82%, respectively. Intra-observer reproducibility in the TM and TFCC was 71% and 76%, respectively. Inter-scanner reproducibility of T2 measurements was 36% in the TM and 85% in the TFCC, respectively.
Conclusion: The assessment of T2 relaxation time measurements of the cartilage of the TM and the TFCC seems to be feasible and reproducible, although the inter-scanner reproducibility of T2 measurements of the TM is suboptimal. Further studies including patients are warranted to prove the utility of this tool.
Keywords: Cartilage; Magnetic resonance imaging; T2 mapping; Trapeziometacarpal joint; Triangular fibrocartilage complex.