Tip Detection Method Using the New IVUS Facilitates the 3-Dimensional Wiring Technique for CTO Intervention

JACC Cardiovasc Interv. 2020 Jan 13;13(1):74-82. doi: 10.1016/j.jcin.2019.07.041. Epub 2019 Dec 11.

Abstract

Objectives: This study assessed the efficacy of the tip detection method during intravascular ultrasound (IVUS)-based 3-dimensional (3D) wiring with a new chronic total occlusion (CTO)-specific IVUS system (AnteOwl IVUS [AO-IVUS]) for CTO percutaneous coronary intervention (PCI).

Background: The study developed angiography-based 3D wiring for CTO-PCI. Previously, the authors produced a short-tip CTO-specific IVUS system (Navifocus WR IVUS [Navi-IVUS]), which has been upgraded into the AO-IVUS system by adding a pullback transducer system for IVUS-based 3D wiring.

Methods: A CTO lesion 20 mm in length composed of 2.5% agar was experimentally inserted into the coronary artery of a beating heart model. The target (a microcatheter with a 0.6-mm lumen) was placed in the distal part of the CTO lesion. IVUS-guided wiring was performed to insert the guidewire into the target using the Navi-IVUS and then using the AO-IVUS 8 times each. In wiring with AO-IVUS, the IVUS-based 3D wiring using the tip detection method was performed. The crossing time and the number of punctures to the target were calculated.

Results: The crossing time was significantly shortened and the number of punctures was significantly reduced in AO-IVUS-based wiring compared with Navi-IVUS-based wiring (median crossing time 80.5 s [interquartile range: 44.0 to 112.3 s] vs. 333.0 s [interquartile range: 88.8 to 790.0 s]; p = 0.036; median 1.0 puncture [interquartile range: 1.0 to 2.0 punctures] vs. 24.0 punctures [interquartile range: 5.8 to 52.5 punctures]; p = 0.001).

Conclusions: The tip detection method enables the authors to easily perform the IVUS-based 3D wiring, and the new CTO IVUS system will facilitate this method in clinical practice.

Keywords: IVUS-based 3D wiring; chronic total occlusion; coronary intervention; tip detection method.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheters*
  • Chronic Disease
  • Coronary Occlusion / diagnostic imaging
  • Coronary Occlusion / physiopathology
  • Coronary Occlusion / therapy*
  • Coronary Vessels / diagnostic imaging*
  • Female
  • Humans
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Predictive Value of Tests
  • Time Factors
  • Ultrasonography, Interventional*