Durability and Clinical Outcomes of Transcatheter Aortic Valve Replacement for Failed Surgical Bioprostheses

Circ Cardiovasc Interv. 2019 Oct;12(10):e008155. doi: 10.1161/CIRCINTERVENTIONS.119.008155. Epub 2019 Oct 14.

Abstract

Background: Valve-in-valve transcatheter aortic valve replacement (TAVR) is an option when a surgical valve demonstrates deterioration and dysfunction. This study reports 3-year results following valve-in-valve with self-expanding TAVR.

Methods: The CoreValve US Expanded Use Study is a prospective, nonrandomized, single-arm study that evaluates safety and effectiveness of TAVR in extreme risk patients with symptomatic failed surgical biologic aortic valves. Study end points include all-cause mortality, need for valve reintervention, hemodynamic changes over time, and quality of life through 3 years. Patients were stratified by presence of preexisting surgical valve prosthesis-patient mismatch.

Results: From March 2013 to May 2015, 226 patients deemed extreme risk (STS-PROM [Society of Thoracic Surgeons Predicted Risk of Mortality] 9.0±7%) had attempted valve-in-valve TAVR. Preexisting surgical valve prosthesis-patient mismatch was present in 47.2% of the cohort. At 3 years, all-cause mortality or major stroke was 28.6%, and 93% of patients were in New York Heart Association I or II heart failure. Valve performance was maintained over 3 years with low valve reintervention rates (4.4%), an improvement in effective orifice area over time and a 2.7% rate of severe structural valve deterioration. Preexisting severe prosthesis-patient mismatch was not associated with 3-year mortality but was associated with significantly less improvement in quality of life at 3-year follow-up (P=0.01).

Conclusions: Self-expanding TAVR in patients with failed surgical bioprostheses at extreme risk for surgery was associated with durable hemodynamics and excellent clinical outcomes. Preexisting surgical valve prosthesis-patient mismatch was not associated with mortality but did limit patient improvement in quality of life over 3-year follow-up.

Clinical trial registration: URL: http://www.clinicaltrials.gov. Unique identifier: NCT01675440.

Keywords: aortic valve; aortic valve stenosis; bioprosthesis; transcatheter aortic valve replacement.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Failure*
  • Quality of Life
  • Recovery of Function
  • Risk Factors
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01675440