Validity and sensitivity to change of laser Doppler imaging as a novel objective outcome measure for cutaneous lupus erythematosus

Lupus. 2019 Oct;28(11):1320-1328. doi: 10.1177/0961203319873977. Epub 2019 Sep 14.

Abstract

Objectives: The objectives of this study were to assess the reliability of a novel objective outcome measure, laser Doppler imaging (LDI), its validity against skin biopsy histology and other clinical instruments, including localized cutaneous lupus disease area and severity index (L-CLASI) and visual analogue scale (VAS) score of photographs, and its responsiveness to clinical change with therapy.

Methods: A prospective observational cohort study was conducted in 30 patients with active cutaneous lupus erythematosus (CLE). At baseline and 3 months, disease activity was assessed using L-CLASI and a high resolution LDI system by two assessors. Skin biopsy was scored as 0 = non-active, 1 = mild activity and 2 = active. Photographs were assessed by two clinicians using 100 mm VAS. Inter-rater reliability was analyzed using Bland-Altman limits of agreement. Correlation between histology and LDI, L-CLASI and VAS and sensitivity to change of LDI with physician subjective assessment of change (PSAC) at 3 months were analyzed using Kendall's tau-a.

Results: Of 30 patients with CLE, 28 (93%) were female, mean (SD) age 48.4 (11.5) y, 25 (83%) were Caucasians, 25 (83%) had concurrent systemic lupus erythematosus and 16 (53%) were smokers. CLE subtypes were acute = 9, subacute = 8 and chronic = 13. Inter-rater agreement for LDI was fair but for VAS score of photographs was poor. In 20 patients with biopsy, correlation with histology was better for LDI (tau-a = 0.53) than L-CLASI (tau-a = 0.26) (difference = 0.27; 90% CI 0.05-0.49) or VAS score of photographs (tau-a = 0.17) (difference = 0.36; 90% CI 0.04-0.68). There was a moderate correlation between PSAC score and change in LDI (tau-a = 0.56; 90% CI 0.38-0.74; p < 0.001, n = 15).

Conclusion: LDI provides a reliable, valid and responsive quantitative measure of inflammation in CLE. It has a better correlation with histology compared to clinical instruments. LDI provides an objective outcome measure for clinical trials.

Keywords: Cutaneous lupus; Doppler; imaging; outcome measure; systemic lupus erythematosus.

Publication types

  • Observational Study
  • Validation Study

MeSH terms

  • Adult
  • Biopsy
  • Cohort Studies
  • Female
  • Humans
  • Laser-Doppler Flowmetry*
  • Lupus Erythematosus, Cutaneous / diagnostic imaging*
  • Lupus Erythematosus, Systemic / diagnostic imaging*
  • Male
  • Middle Aged
  • Observer Variation
  • Outcome Assessment, Health Care*
  • Prospective Studies
  • Reproducibility of Results
  • Sensitivity and Specificity