Incidence, predictors and cerebrovascular consequences of leaflet thrombosis after transcatheter aortic valve implantation: a systematic review and meta-analysis

Eur J Cardiothorac Surg. 2019 Sep 1;56(3):488-494. doi: 10.1093/ejcts/ezz099.

Abstract

Objectives: We examined the incidence, the impact of subsequent cerebrovascular events and the clinical or procedural predictors of leaflet thrombosis (LT) in patients undergoing transcatheter aortic valve implantation (TAVI).

Methods: MEDLINE/PubMed was systematically screened for studies reporting on LT in TAVI patients. Incidence [both clinical and subclinical, i.e. detected with computed tomography (CT)] of LT was the primary end point of the study. Predictors of LT evaluated at multivariable analysis and impact of LT on stroke were the secondary ones.

Results: Eighteen studies encompassing 11 124 patients evaluating incidence of LT were included. Pooled incidence of LT was 0.43% per month [5.16% per year, 95% confidence interval (CI) 0.21-0.72, I2 = 98%]. Pooled incidence of subclinical LT was 1.36% per month (16.32% per year, 95% CI 0.71-2.19, I2 = 94%). Clinical LT was less frequent (0.04% per month, 0.48% per year, 95% CI 0.00-0.19, I2 = 93%). LT increased the risk of stroke [odds ratio (OR) 4.21, 95% CI 1.27-13.98], and was more frequent in patients with a valve diameter of 28-mm (OR 2.89: 1.55-5.8), for balloon-expandable (OR 8: 2.1-9.7) or after valve-in-valve procedures (OR 17.1: 3.1-84.9). Oral anticoagulation therapy reduced the risk of LT (OR 0.43, 95% CI: 0.22-0.84, I2 = 64%), as well as the mean transvalvular gradient.

Conclusions: LT represents an infrequent event after TAVI, despite increasing risk of stroke. Given its full reversal with warfarin, in high-risk patients (those with valve-in-valve procedures, balloon expandable or large-sized devices), a protocol which includes a control CT appears reasonable.

Keywords: Cerebrovascular events; Leaflet thrombosis; Stroke; Transcatheter aortic valve implantation.

Publication types

  • Meta-Analysis
  • Systematic Review

MeSH terms

  • Anticoagulants / therapeutic use
  • Aortic Valve Stenosis / surgery
  • Computed Tomography Angiography
  • Female
  • Heart Valve Prosthesis / adverse effects
  • Humans
  • Male
  • Observational Studies as Topic
  • Randomized Controlled Trials as Topic
  • Risk Factors
  • Stroke / chemically induced*
  • Stroke / prevention & control
  • Thrombosis / etiology*
  • Transcatheter Aortic Valve Replacement / adverse effects*

Substances

  • Anticoagulants