Accuracy of the point-of-care glucose meter for use in calves

J Vet Med Sci. 2019 Sep 18;81(9):1301-1304. doi: 10.1292/jvms.19-0272. Epub 2019 Jul 15.

Abstract

The aim of this study was to evaluate the accuracy and precision of portable blood glucose meters, such as i-STAT 1 and Precision Xceed, for use in calves. Whole blood and plasma samples were obtained from eleven calves that received 2.5 or 5.0% dextrose-containing polyelectrolyte isotonic solutions. Measurements using the i-STAT 1 (r2=0.99, P<0.0001) and Precision Xceed (r2=0.96, P<0.0001) were well correlated with those by the hexokinase method, which is the gold standard. Although the accuracy of i-STAT 1 was equivalent to that of the hexokinase method, there was an autocorrelation in the residuals between the results from the Precision Xceed and the hexokinase method. Thus, the i-STAT 1 can be used to measure the blood glucose concentration in cattle.

Keywords: accuracy; cattle; cow-side test; glucose; portable blood glucose meter.